MicroLog Plus:

What external sensors can be plugged into MicroLog?

MicroLog will work with the following sensors:

Back to MicroLog plus FAQs

 

When logging with external sensor, Et appears on the MicroLog display. What does it mean?

MicroLog displays readings of internal sensors and the external NTC temperature sensor. When working with any other external sensor you can display readings only by downloading the data to the software. The MicroLog display shows Et while taking measurements of these sensors.

 

Back to MicroLog plus FAQs

 

For how long will the MicroLog's battery last?

The MicroLog battery's maximum lifespan is approximately 1.5 years. If you use high sampling rates, the MicroLog battery's lifespan will be considerably shorter (see figure below):

Back to MicroLog plus FAQs

 

Can data be downloaded to a computer using the IR port?

No, data can be downloaded either by cable communication or by wireless RF transmission

 

Back to MicroLog plus FAQs

 

How can I know in what mode the MicroLog is working?

Press the two MicroLog's button simultaneously for one second:

The MicroLog will display its status:

Run
MicroLog is recording data. It will stop recording automatically when its memory is full (16,000 recordings)

 

Cyclic Run
Same as Run, but MicroLog will record over the old data when the memory is full, beginning with the earliest data recording

 

Back to MicroLog plus FAQs

 

How often should I calibrate the MicroLog?

The MicroLog should be checked for calibration once a year. We recommend to send the MicroLog to a certified calibration laboratory or to return it to fourtec for factory calibration.

 

Back to MicroLog plus FAQs

 

Does the calibration procedure update both the MicroLog and the software?

The humidity calibration procedure updates both the MicroLog and the software. You will see correct readings both in MicroLog’s display and in the software.

The temperature calibration, however, will update the software only

 

Back to MicroLog plus FAQs

 

When calibrating the humidity sensor I clicked the Default button and now I see strange humidity readings

Repeat the calibration procedure. This time be careful not to touch the humidity sensor with your fingers when clicking the Default button

 

Back to MicroLog plus FAQs

 

I'm using a Cradle and a receiver but I get poor RF reception

The usable transmission range is 300m line of sight. physical obstacles like walls and buildings may reduce the range considerably.

You can use fourtec's RF repeater to double the usable range or to overcome line of sight problems. In this case place the repeater so it can "see" both the Cradle and the receiver

 

Back to MicroLog plus FAQs